| Sign In | Join Free | My futurenowinc.com |
|
| Categories | Battery Capacity Discharge Tester |
|---|---|
| Brand Name: | Neware |
| Model Number: | CT-4008T-5V12A-S1 |
| Certification: | CE |
| Place of Origin: | China |
| MOQ: | 1 Unit |
| Price: | USD negotiation per Unit |
| Payment Terms: | T/T |
| Supply Ability: | 100 Units Per Month |
| Delivery Time: | 8 work days |
| Packaging Details: | Carton |
| Voltage Range: | 1.5V - 5V |
| Channel Current Range: | 5mA - 6A - 12A |
| Min Current record condition: | 2mA - 12mA - 24mA |
| Noise: | Less than 85dB |
| Data Output: | EXCEL , TXT |
| Database: | MYSQL |
| Company Info. |
| Neware Technology Limited |
| View Contact Details |
| Product List |
Neware 5V12A Battery Capacity Discharge Test System Pulse Discharge Analyzer
Recycle Battery Function
1 . If you have recovery battery pack , like Toyota Prius battery pack - generally , each piece has 6 single NiMH cell , then several pieces batteries assembed in one pack , and you need to check each single cell for capacity , cyclie life , resistance , etc .
2 . Then assemble the battery cell after sorting by capacity , resistance . You will get the new battery pack . And last you will test the battery pack to check the real capacity and other function parameter .
Voltage and Current - the Very Essential
1 . The accuracies of Voltage and / or current may fundamentally affect the results come from a test . Thats why accuracies of voltage and current are so important to a battery tester .
2 . Neware battery testers provide good accuracies both for voltage and current . This ensures you get the most reliable data from your tests .
3 . A measurement system can be accurate but not precise , precise but not accurate , neither , or both . Neware battery testers provide good prrecision meanwhile .
4 . To get an overall performance of a measurement system , Measurement system analysis MSA is usually introduced for the study .
20ms Response Time
1 . This fast response time will meet rapidly changing power
requirements
2 . Flexible Integration of Hardware and Data Management
3 . BTS 4000 functions also as the core part of auto formation and
grading system .
4 . With its structural data system , testing data can be analysed , summarized , managed and utilized by our extended program or customers own data and logic server .
| Input Power | 425W | |
| Channel Features | Constant Current Source and Constant Voltage Source with Independent pairs of Closed - loop Structure | |
| Channel Control Mode | Independent Control | |
| Voltage | Per Channel Voltage Range | 1.5V - 5V |
| Accuracy | 0.05 percent of FS | |
| Current | Per Channel Current Range | Range1 , 5mA - 1A Range2 , 1A - 6A Range3 , 6A - 12A |
| Accuracy | 0.05 percent of FS | |
| Stability | 0.05 percent of FS | |
| Data Recording | Record Condition | Min Time - 100mS |
| Min Voltage - 10mV | ||
| Min Current - Range 1 , 2mA Range 2 , 12mA Range 3 , 24mA | ||
| Record Frequency | 10Hz | |
| Charge | Charge Mode | CCCharge , CVCharge , CCCVCharge , CPCharge |
| End Conditon | Voltage , Current , Relative Time , Capacity , -deltaV | |
| Discharge | Discharge Mode | CCD , CPD , CRD |
| End Condition | Voltage , Current , Relative Time , Capacity | |
| Pulse | charge | CCC , CPC |
| Discharge | CCD , CPD | |
| Min Pulse Width | 500ms | |
| Automated Switch | Automated swith from charge to discharge for each pulse | |
| Cycle | Loop Measure Range | 65535 times |
| Max Steps Per Loop | 254 | |
| Nested Loop | Nested loop Function , Max Support 3 Layers | |
| Protection | Software Protection | Power - down Data Protection |
| Off - line Testing Function | ||
| Setting Protection Condition,Setting Parameter - Low Voltage limit , Upper Voltage limit , Low Current limit , Upper Current limit , Delay Time | ||

|