| Brand Name: | HongCe |
| Model Number: | HCAO1000 |
| Certification: | Calibration certificate(cost additional) |
| Place of Origin: | China |
| MOQ: | 1 set |
| Price: | Negotiation |
| Payment Terms: | L/C, T/T |
| Supply Ability: | 10 sets per month |
| Delivery Time: | 30 days |
| Packaging Details: | Plywood |
| Product name: | Aging Oven |
| Warranty: | 1 Year |
| Power: | AC380(±10%)V/50HZ |
| Standard: | IEC 60811-1-2, ISO 188 |
| Application: | heat resistance test |
| Temperature range: | RT+10℃~200℃ |
| Temperature fluctuation: | ≤±0.5℃ |
| Air exchange frequency: | 1~100times /h (Adjustable) |
| Company Info. |
| Guangzhou HongCe Equipment Co., Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |
IEC 60811-1-2 / ISO 188 IEC Test Equipment Aging oven 10℃ - 200℃
Product information:
It is conforms to ISO 188 ‘Rubber, vulcanized or thermoplastic — Accelerated ageing and heat resistance tests’, clause 8.1 of IEC 60811-1-2 ‘Common test methods for insulating and sheathing materials of electric and optical cables — Part 1-2: General application — Thermal ageing methods’ and etc. .
The equipment is suitable for the heat resistance test of electrical insulating materials, the performance and parameters of electronic parts and plastic products under high temperature and atmospheric pressure
Parameters:
| Chamber inner size | D1000mm*W1000mm*H1000mm |
| Equipment overall size | D1650mm*W1420mm*H2100mm |
| Temperature range | RT+10℃~200℃ |
| Temperature fluctuation | ≤±0.5℃ |
| Temperature uniformity | ≤±2℃ |
| Resolution | ±0.1℃ |
| Air exchange frequency | 1~100times /h(Adjustable ) |
| Time setting range | 0~9999H |
| Ventilation time | 1min~99 H |
| Cycle mode | fan forced circulation mode |
| Turntable rotating speed | 1~5r/min (Adjustable ) |
| Turntable diameter | 600mm, can hung 24 samples holders |
| Air flow rate | 0.5-1.5m/s |
| Power supply | AC380(±10%)V/50HZ |
Heating system:
1. Far-infrared nickel alloy high-speed heating electric heater;
2. High temperature, ventilation completely independent system;
3. Temperature control output power is calculated by microcomputer to achieve high efficiency and high efficiency.
Control System:
1. Digital display LED touch button type PID microcomputer SSR output temperature controller
2. Accuracy: 0.1°C (display range)
3. Resolution: + 0.1 C;
4. Temperature sensor: PT100 platinum resistance thermometer;
5. Control method: thermal balance temperature adjustment method;
6. Temperature control adopts P.I.D+S.S.R system co-channel coordination control;
7. With automatic calculation function, it can reduce inconvenience caused by human setting.
Ventilation Control:
1. Natural ventilation: ventilation times: 1 to 20 times/hour
2. Natural ventilation + forced ventilation: 1-100 times / hour
3. Forced air change control: Use air pump to inflate, by adjusting the flow rate of the air flow meter
Protection System:
Over-temperature protection, Timing protection, Overload protection, Fan overheat protection
Equipment Use Environment:
1. Ambient temperature: 5 ~ + 28 ℃ (Average temperature within 24 hours≤28℃) ;
2. Environmental humidity: ≤85%R.H;


|