Repeatability: ≤ ±0.5% F.S.
Drift: ≤ ±1% F.S./8h
VPR Removal Efficiency: >99.0% (C40)
Minimum Particle Size Limit: 23 nm or 10 nm
Measurement Principle: Electrical charge counting method
Measuring Range: 600 - 1.3×10⁹ #/cm³
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

KGD-WQ-504 Portable Emissions Tester (PEMS)

KGD-WQ-504 Portable Emission Testing System (PEMS)

Product Introduction

The KGD-WQ-504 Portable Emission Testing System is specifically designed for in - use vehicle on - road emission tests. It can continuously and real - time measure or collect the volume concentrations of air pollutants such as CO, CO₂, NO, NO₂, NOₓ, O₂, and THC, as well as particulate matter number concentration, exhaust gas flow, global positioning data, meteorological data, OBD parameters, etc. It has functions such as test process guidance, automatic calculation of total emissions, and export of test reports, and can meet the certification tests, in - use compliance tests, and related research and development test requirements of global vehicle on - road emission regulations.

Product Features

  • Core sensing technology under independent control
  • One - stop service throughout the whole life cycle
  • Small in size, light in weight, and low in energy consumption
  • Easy to install and use, with a lithium - ion battery life of over 8 hours
  • Modular design, plug - and - play
  • Flexibly meet the needs of various regulatory and R & D tests
  • High - standard waterproof and shockproof
  • Maintain stable test performance under various harsh working conditions

Technical Parameters


Host SystemHost SystemHost SystemHost SystemHost System
Gas ComponentCOCO₂NONO₂
Measurement PrincipleNDIRNDIRNDUVNDUV
Measuring Range0 - 0.5 - 5%0 - 5 - 20%0 - 3000 ppm0 - 1000 ppm
Response TimeT10 - 90 < 2.5s
Accuracy≤±2.0%RS or ≤±0.3%FS
LinearityDetermination coefficient: ≥ 0.998; Standard deviation: ≤ 1% F.S.; Slope: 0.99 - 1.01; Intercept coefficient: ≤ 0.5% F.S.
Repeatability≤ ±0.5% F.S.
Drift≤ ±1% F.S./8h
Operating EnvironmentOperating temperature: -10 - 45°C; Relative humidity: ≤ 90%RH; Altitude: 0 - 3000m
Exhaust Gas FlowExhaust Gas Flow
Temperature Range-5 - 500°C
Exhaust Gas Temperature Measurement Accuracy≤ ±1% reading or ± 2°C
Exhaust Gas Flow Measurement Range10 - 600kg/h; 100 - 2000kg/h
Exhaust Gas Flow Measurement Accuracy≤ ±2.0% reading or ≤ ±0.5% full - scale value
Exhaust Gas Flow Measurement LinearityDetermination coefficient: ≥ 0.990; Standard deviation: ≤ 1% full - scale value; Slope: 0.99 - 1.01; Intercept coefficient: ≤ 1% full - scale value
Response Time< 1s
THC/CH₄ AnalyzerTHC/CH₄ Analyzer
Gas ComponentTHCTHCCH₄CH₄
Measurement PrincipleHFIDHFIDNMC - FIDNMC - FID
Minimum Range0 - 100 ppmC0 - 100 ppmC0 - 100 ppmC0 - 100 ppmC
Maximum Range0 - 10000 ppmC0 - 10000 ppmC0 - 3000 ppmC0 - 3000 ppmC
Response TimeT10 - 90 < 2.5s
Accuracy≤ ±2.0% MRS or ≤ ±0.3% F.S.
LinearityDetermination coefficient: ≥ 0.998; Standard deviation: ≤ 1% F.S.; Slope: 0.99 - 1.01; Intercept coefficient: ≤ 0.5% F.S.
Repeatability≤ ±0.5% F.S.
Drift≤ ±1% F.S./8h
PN AnalyzerPN Analyzer
Minimum Particle Size Limit23 nm or 10 nm
Measurement PrincipleElectrical charge counting method
Measuring Range600 - 1.3×10⁹ #/cm³
Measurement Accuracy≤ ±10%
VPR Removal Efficiency>99.0% (C40)
Response TimeT90 < 5s
NH₃/N₂O AnalyzerNH₃/N₂O Analyzer
Gas ComponentNH₃NH₃N₂ON₂O
Measuring Range0 - 1000 ppm
Response TimeT10 - 90 < 2.5s
Accuracy≤ ±2.0% MRS or ≤ ±0.3% F.S.
LinearityDetermination coefficient: ≥ 0.998; Standard deviation: ≤ 1% F.S.; Slope: 0.99 - 1.01; Intercept coefficient: ≤ 0.5% F.S.
Repeatability≤ ±0.5% F.S.
Drift≤ ±1% F.S./8h