| Sign In | Join Free | My futurenowinc.com |
|
| Categories | Online Water Quality Analyzer |
|---|---|
| Brand Name: | JUGE |
| Model Number: | AG9010Zn |
| Certification: | IOS9001, CE |
| Place of Origin: | Nanjing, China |
| MOQ: | 1 Unit |
| Price: | USD2000-7000 |
| Packaging Details: | Neutral carton |
| Delivery Time: | 5-8 working days |
| Payment Terms: | Online order,T/T |
| Supply Ability: | 100 Unit/Week |
| Test Method: | Zincon Colorimetric Method |
| Measuring Range: | 0~30 mg/L (segmented measurement, expandable) |
| Detection Limit: | ≤0.02 mg/L |
| Resolution: | 0.001 mg/L |
| Accuracy: | ±5% |
| Power Supply: | AC 220±10% V |
| Dimensions: | 520H × 370W × 265D (mm) |
| Weight: | 40kg |
| Company Info. |
| Nanjing Juge Environmental Technology Co., Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |
Zinc Online Water Quality Analyzer
>>Overview
>>Principle
The product employs the spectrophotometric method for zinc detection. After mixing the water sample with a conditioning agent, all forms of zinc are converted to zinc ions (Zn²⁺). In an alkaline environment and in the presence of a sensitizer, Zn²⁺ reacts with an indicator to form a colored complex. The analyzer detects this color intensity change, converts it into zinc concentration values, and outputs the results. The amount of colored complex generated corresponds to the total zinc concentration.
>>Technical Parameter
| Test Method | Zincon Colorimetric Method |
| Measuring Range | 0~30mg/L (segmented measurement, expandable) |
| Detection Limit | ≤0.02mg/L |
| Resolution | 0.001mg/L |
| Accuracy | ±5% |
| Repeatability | ≤5% |
Drift (24h) | Zero drift: ±5% Span drift: ±10% |
| Measurement Time | <30min (digestion time adjustable) |
| Measurement Modes | Adjustable time interval, fixed time point, or triggered measurement |
Calibration Modes | Automatic calibration (1~99 days adjustable) Manual calibration based on actual water samples |
Maintenance Cycle |
|
| Human-Machine Interface | Touchscreen display and command input |
Self-Diagnosis & Protection | Self-diagnoses operational status; retains data during anomalies/power outages Automatically clears residual reactants and resumes operation after reset/power recovery |
| Data Storage | 5-year data capacity |
| Input Interface | Digital input |
Output Interfaces | 1× RS232, 1× RS485 2× 4~20mA analog outputs |
Operating Environment | Indoor use; recommended temperature: 5~28℃, humidity ≤90% (non-condensing) |
Power Supply | Voltage: AC 220±10% V Frequency: 50±0.5 Hz Power: ≤150W (excluding sampling pump) |
Dimensions | 520H × 370W × 265D (mm) |
|