| Basic parameters | |
| Measuring medium | Liquids, fluids, gases (compatible with contact materials) |
| Measuring range | -0.1~0~60MPa (see range selection table for details) |
| Type of pressure | Gauge pressure, absolute pressure, negative pressure |
| Temperature compensation | -10~70℃ |
| Accuracy class | 0.25%FS,0.5%FS |
| Frequency of response | 5,30Hajj100Hajj |
| Stable performance | ±0.1% FS/year |
| Temperature drift | ±0.01%FS/°C (within temperature compensation)±0.05%FS/°C (outside
temperature compensation) |
| Overall weight | No display ≈ 190g with display ≈ 270g |
| Ingress protection | IP65 (no display) IP54 (display) Note: The above protection level
refers to the completion of the electrical connection |
| Output power | |
| Output signal | 4~20mA,RS485,0~10VDC,0~5VDC,1~5VDC,0.5~2.5VDC |
| Power supply range | 9~28VDC,15~28VDC,3~5VDC,15~24VDC |
| Power range | |
| Voltage type | ≤0.02Us(W),≤0.008Us(W),≤0.001Us(W) |
| Current type | ≤0.02Us(W) |
| Digital | ≤0.02Us(W),≤0.015Us(W) |
| Load characteristics | |
| Voltage type | ≥10kΩ |
| Current type | No display load ≤{(Us-7.5)÷0.02}Ω With display load≤{(Us-12)÷0.02}Ω (Us=supply voltage) |
| Environmental conditions | |
| Medium temperature | -40~85℃ |
| Ambient temperature | -40~85℃ |
| Ambient humidity | 0%~95%RH(无冷凝无结露) |
| Overload capacity | |
| Overload pressure | 300%FS,200%FS,12kPa,50kPa,300kPa,480kPa,750kPa,800kPa,1.2MPa,2MPa |
| Burst pressure | 600%FS,500%FS,400%FS,20kPa,200kPa,600kPa,900kPa,1.25MPa,2MPa,3MPa,5MPa,2MPa,3MPa,5MPa |
| Overall material | |
| diaphragm | 316L stainless steel (in contact with the medium to be measured) |
| Process connections | 316 stainless steel (in contact with the medium to be measured) |
| enclosure | 304 stainless steel |
| O-rings | Nitrile rubber (contact with the medium to be measured) |
| ED CIRCLE | Nitrile rubber (contact with the medium to be measured) |
| Displays the shell | ABS Engineering Plastics (LED Display) |
| Hersmann connectors | ABS engineering plastics |
| Mechanical stability | |
| Vibration resistance | 10g(20… 2000Hz) in accordance with IEC60068-2-6 standard |
| Impact Resistance: | 500g/ms meets IEC60068-2-27 standard |
| Electrical protection | |
| Short-circuit protection | eternity |
| Anti-polarity protection | No damage, but no work |
| Insulation resistance | ≥100MΩ500VDC |
| Dielectric Strength: | 500VAC |
| Output limits | |
| The output is minimal | 4mA,0,0.001VDC,0.05VDC |
| The output is maximum | 20mA,2000,10VDC,5VDC |